MCUSTOM_CHOICEVARIANTS
gigazyme®
Enzymatic cleaner for the manual reprocessing of medical devices, such as e.g. flexible endoscopes and surgical instruments. instruments.Enzymatic cleaner for manual reprocessing of medical devices, such as flexible endoscopes and surgical instruments.
- First-class cleaning power due to 3 enzymes
- Excellent material compatibility due to neutral pH-value
- Low foaming due to non-ionic surfactants
Areas of application
gigazyme® is an enzymatic cleaning concentrate for the manual reprocessing of medical devices, such as flexible endoscopes and surgical instruments. Enzyme & surfactant power: The optimum combination of high-quality enzymes (protease = breaks down proteins, lipase = dissolves fatty soiling, amylase = removes polysaccharides) and an innovative surfactant system results in excellent cleaning of all types of thermostable and thermolabile medical devices with outstanding material compatibility.
Instructions for use
gigazyme® can be used in immersion baths or ultrasonic baths.
Dosage:
gigazyme® is a concentrate and is diluted with water to the desired application concentration. Water temperature < 45 °C to avoid protein denaturation. The dosage depends on the type of reprocessing and the degree of soiling of the instruments. The following parameters are recommended when using gigazyme®:
For normal soiling: 0.5 - 2 % (1 litre of a 1 % solution corresponds to 990 ml water and 10 ml gigazyme®).
For heavy soiling, up to 5 %. Place the instruments to be cleaned in the working solution. Ensure complete wetting, also for hollow instruments, and allow to act until the instruments are visually clean. After cleaning, rinse/flush the medical devices thoroughly with water of at least drinking water quality to completely remove residues of the solution.
Standing time:
It is recommended to renew the working solution at least every working day or immediately in case of visible contamination.
Please observe the reprocessing recommendations of the instrument manufacturers.
- Signal word: Warning
Danger warnings
Danger warnings
H319 - Causes serious eye irritation.Safety Instructions
P280 - Wear protective eye protection/face protection.
P305 + P351 + P338 - IF IN EYES: Rinse cautiously with water for several minutes. Remove contact lenses, if present and easy to do. Continue rinsing.
P337 + P313 - If eye irritation persists: Get medical advice/attention.